Tripp Lite Power Supply 700 HG User Manual

Owner's Manual Addendum to #93-1424  
SmartPro® 700 HG  
Hospital-Grade, Intelligent  
1111 W. 35th Street  
Chicago, IL 60609 USA  
Customer Support: (773) 869-1234  
Line-Interactive UPS System (120V)  
ESPAÑOL: p. 1-2  
FRANÇAIS: p. 2  
The SmartPro 700 HG systems described in this addendum has the same features  
and controls as the SmartPro 700 system described in the accompanying owner's  
manual. However, it is a true hospital-grade system that meets the UL 544 standard  
for medical and dental equipment.  
Hospital Grade Specifications: The SmartPro 700 HG exceeds the requirements of  
Underwriter's Laboratories Standard 544. UL 544 describes the low-leakage electrical  
characteristics required of medical and dental equipment used in patient care areas.  
It specifies a maximum leakage current of 300 microamps. The SmartPro 700 HG  
incorporates on-line transformer isolation, and is designed to exhibit less than 100  
microamps leakage to ground. The SmartPro 700 HG's inverter, which incorporates a  
double-pole line disconnect, operates during AC power failures and extreme voltage  
events and is designed to maintain the low leakage level specified by UL 544.  
Warning: Tripp Lite does not recommend the use of any of its UPS Systems in any  
life support applications in which a malfunction or failure of a Tripp Lite UPS System  
could cause failure or significantly alter the performance of a life-support device.  
Specifications  
SMART700HG  
Output Capacity (VA/Watts):  
Battery Runtime  
(Half Load/Full Load) Minutes:  
Battery Recharge Time:  
Approvals:  
700/450  
20/6  
2-4 hrs.  
UL, cUL, NOM  
Input Voltage/Frequency (120V/60 Hz); On-Line Input Voltage Range (79 - 143 volts); Voltage-  
Regulated Output Voltage Range (120V 9%); On Battery Output Voltage Range (120V 5%);  
Output Waveform Line Mode (filtered sinewave); Output Waveform Battery Mode (PWM sine wave);  
AC Surge Suppression (exceeds IEEE 587 Cat. A & B standards); AC Noise Attenuation (>40 dB);  
AC Protection Modes (H to N, H to G, N to G).  
ESPAÑOL  
El No-break / Sistema UPS SmartPro 700 HG especificado en este documento incluye  
las mismas características y controles del sistema SmartPro 700 descrito en al  
manual de operación adjunto. No obstante, esta unidad es de calidad para hospitales  
ya que cumple con las normas UL 544 para equipos de uso médico y dental.  
Especificaciones de Calidad para Hospitales: El SmartPro 700 HG excede los  
requisitos de la Norma 544 de Underwriter’s Laboratories. Dicha norma especifica  
las características de baja corriente de fuga requeridas para equipos de uso médico  
o dental en situaciones de cuidados de pacientes. Esta norma especifica que la  
corriente máxima de fuga no debe exceder 300 microamperios. El SmartPro 700 HG  
incorpora el aislamiento en línea del transformador y ha sido diseñado para exhibir  
corriente de fuga menor a 100 microamperios. El inversor del SmartPro 700 HG, el  
cual incorpora una línea de polaridad doble, opera durante fallas en el suministro de  
 

Thermador Cooktop CIT304 User Manual
Timex Watch W254 User Manual
Toastmaster Fryer 1427 User Manual
Tokina Camcorder Accessories MVH 2D User Manual
Toshiba Camcorder B10 User Manual
Tripp Lite Computer Accessories NB1000GY User Manual
Troy Bilt Trimmer TB32 EC User Manual
Uniden Cordless Telephone EXI7960 User Manual
Vintage Gas Grill VBQ 56 User Manual
Vizio Flat Panel Television VX52L User Manual