Omron Respiratory Product NE U07 User Manual

NE U07  
Instruction manual  
Page 2  
Gebrauchsanweisung  
Seite 20  
Mode demploi  
Page 38  
Istruzioni per luso  
Pagina 56  
Instrucciones de uso  
Página 74  
Inhalationsgerät – Nebuliser  
N E U07 5.1 CE  
 
English  
General Inform ation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3  
Com ponents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4  
Safety precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6  
Inform ation on inhalation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8  
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9  
Start the treatm ent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10  
End of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11  
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11  
Possible problem s and how to solve them . . . . . . . . . . . . . . . .14  
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15  
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16  
Guarantee card . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17  
The OMRON N E U07 is a nebuliser for the inhalation of m edi-  
cal aerosols to treat asthm a, chronic bronchitis and other respi-  
ratory tract disorders. This nebuliser is suitable for com m only  
used inhalation solutions. Use only the m edication prescribed  
by your doctor. Type and duration of the inhalation m ust be  
carried out according to the recom m endations of your doctor.  
Table of  
contents  
General  
Information  
The OMRON N E U07 produces a very low particle size  
distribution and therefore a m edication vapour that will reach  
effectively and reliably to the upper and lower respiratory  
passages.  
The OMRON N E U07 can be used autom atically with constant  
aerosol-delivery function or interm ittently by an interrupt  
switch.  
Thank you for purchasing an OMRON product. You can be sure  
you have m ade a wise choice. By buying the OMRON N E U07  
nebuliser, you have chosen a high-quality, innovative instru-  
m ent. During its developm ent, particular attention was given to  
reliability and easy, convenient handling. Before using for the  
first tim e, please read through this m anual carefully. If you  
should still have any questions regarding its use, please contact  
the OMRON Custom er Service at the address on the packaging  
and/or attached literature. They will be pleased to help you.  
Dear  
Customer  
The OMRON N E U07 is suitable for inhalation both at hom e  
and in hospital or a doctors practice.  
The OMRON N E U07 nebuliser corresponds to the appropriate  
CE regulations.  
Best wishes for a healthy future.  
Yours sincerely,  
OMRON HEALTHCARE Europe  
2
3
 
English  
1
2
Components  
of the  
14  
OMRON  
NE U07  
main unit  
15  
3
4
NE U07  
5
6
18  
19  
9
7
10  
20  
21  
16  
11  
12  
8
13  
22  
23  
17  
1 Mouthpiece  
2 Aerosol rate adjust knob  
3 Cover  
NE U07  
4 Casing  
5 N ebuliser lam p  
6 Battery discharge lam p  
(only if rechargeable batteries are used)  
7 Power cord  
14 Aerosol delivery rate adjust knob  
15 Air filter  
16 Fan  
17 Transducer  
8 Plug  
18 N ebuliser guide  
19 Medication cup  
20 Medication cup holder  
21 Ventilation hole  
22 Water tank  
9 Manual operation button  
10 Continuous operation button (  
11 Unlock button  
), OFF/ON  
12 AC adapter  
13 Bottom housing  
23 Water level detector  
4
5
 
English  
Please read the following instructions carefully, as they contain  
im portant inform ation concerning the setting up, use and  
m aintenance of your new nebuliser.  
When using any electrical appliance certain im portant safety  
m easures m ust always be observed, which include the follo-  
wing:  
Safety  
Precautions  
Keep this m anual for future reference.  
The unit is not protected against water penetration, so:  
N ever subm erse the nebuliser in water.  
As soon as you have unpacked the nebuliser, check it for  
visible dam age or defects. In particular check for cracks in  
the plastic housing, which could expose electrical com po-  
nents. If this is the case or if you are in doubt, contact an  
authorized dealer or the OMRON Custom er Service directly.  
Only use the nebuliser for its intended purpose, that m eans  
for inhalation therapy. Any other form of use constitutes an  
im proper use and is therefore dangerous. The m anufacturer  
cannot be held liable for any dam age caused by im proper or  
incorrect use.  
N ever touch the unit when your hands are wet or m oist.  
If it falls into a filled bathtub or sink, first disconnect the  
power supply plug from the wall socket, before attem pting  
to retrieve it. Im m ediately contact an authorized dealer or  
the OMRON Custom er Service and have the unit inspected.  
Do not use the unit before this inspection and any neces-  
sary repairs have been carried out.  
Do not allow unsupervised children or infirm persons to use  
the unit.  
N ever open the m ain unit.  
Do not disconnect the unit by sim ply pulling out the plug  
from the wall socket. Always switch the unit off first.  
Do not leave the nebuliser running unattended. Always  
switch the unit off and then unplug it when it is in use.  
Do not leave the unit outdoors or in a hum id environm ent.  
N ever store the unit in direct sunlight.  
Do not operate the unit in the presence of anaesthetic m ix-  
tures which are inflam m able when they com e into contact  
with air, oxygen or nitrous oxide.  
If you discover that the nebuliser is dam aged or does not  
operate properly, switch it off im m ediately and stop using  
it. Any repairs m ust be carried out by an authorized dealer  
using only original OMRON spare parts. Any part of the  
unit m ay only be repaired or replaced by the m anufacturer  
or the OMRON Custom er Service.  
Keep the power cable away from heated or hot surfaces.  
The power cable should always be fully unwound in order  
to prevent dangerous overheating.  
Do not use the nebuliser while you are in a bath or shower.  
Som e parts of the unit are so sm all that they could be swal-  
lowed by sm all children. Do not let unattended children  
have access to the unit.  
Make sure that the electrical rating shown on the rating  
plate on the bottom of the unit corresponds to your m ains  
voltage and frequency before plugging in the nebuliser.  
If the power plug provided with the nebuliser does not fit  
your wall socket, have the plug changed by an electrical  
appliance dealer. Do not use an adapter. If you use an exten-  
sion cable m ains adapter m ake sure not to connect too  
m any appliances sim ultaneously so that the m axim um load  
is not exceeded.  
Always disconnect the nebuliser from the power supply  
before com m encing cleaning.  
This appliance is designed for short term operation only.  
The unit m ustnt be operated for longer than 30 m inutes.  
Please allow 30 m inutes between two operations.  
If the power cable or the power supply plug becom e dam a-  
ged, consult an authorized dealer / service agent or contact  
the OMRON Custom er Service. N ever operate the unit  
with a dam aged power cable or plug.  
6
7
 
English  
N ever subm erge or wet the AC power unit or cord. Doing so  
could result in electrical shock.  
Filling cooling water  
Preparation  
Before filling cooling water or medication, be sure the power is  
OFF ( ). (Nebuliser lamp is OFF and the power cord is unplugged).  
N ever operate the unit with the water tank em pty. Use only  
water in the tank. Solutions other than water m ay cause da-  
m age to the transducer.  
Remove the cover by slightly pushing the rear part of the cover to  
move it upwards.  
Lift the medication cup holder by slightly pushing the front panel  
of the tank holder and lifting it up.  
Change the water in the water tank after each use.  
NE-U07  
Warning: N ever rem ove the cover during operation – severe  
injury could occur.  
Filling Medication  
Place the nebuliser on a table or another flat and stable surface.  
Fill the water tank ( with tap water or distilled water) to the  
correct level. (In between the two lines).  
Correct  
water level  
NE-U07  
Caution:  
1. N ever fill the water tank with solutions other than water.  
Dam age to the transducer could occur.  
2. Be sure to fill the water tank to the proper level. The unit  
will not operate if the water level is low.  
When inhaling you should sit upright and relaxed. You should  
sit at a table and not in an arm chair as otherwise you will con-  
strict your respiratory passages and im pair the effect of inhala-  
tion. If you m ust be treated in bed, support yourself with  
pillows so that you will sit as upright as possible.  
Information  
on inhalation  
Place m edication directly into the m edication cup.  
N ever use m ore than 10 m l of solution.  
Be sure all parts are clean before operating.  
One treatm ent should not last longer than 15 m inutes. Consult  
your doctor concerning the appropriate inhalation tim e.  
Always be calm and relaxed when you inhale. Allow tim e for  
the inhalation to take effect after you have finished using the  
nebuliser.  
The m edication cup has a 10 m l capacity. N ever overfill.  
Breathe in slowly and deeply so that the m edication can reach  
deep into the fine bronchial tubes. Hold your breath for a short  
time, then breathe out slowly. Do not try to breathe too quickly.  
Take a pause when you feel like having it.  
Before using the nebuliser for the first tim e it m ust be cleaned  
according to the cleaning instructions.  
8
9
 
English  
Aerosol output is adjustable by rotating the delivery rate  
adjust knob placed on the cover.  
Assem bling  
Rotate the knob counterclockwise to increase the rate, or  
clockwise to reduce the rate. Adjust it to your requirem ents.  
Mount the medication cup holder and the cover onto the main  
unit.  
NE-U07  
Sit relaxed at a table and breathe in the aerosol delivered  
through the m outhpiece or air tube (optional part).  
NE-U07  
Couple the main unit with the AC adaptor. (Fit them by pressing  
until it clicks).  
Press the continuous operation button (  
(N ebuliser lam p goes out).  
) after use.  
End of  
operation  
Connect the plug to a wall-plug (AC 220V – 230V~).  
Start the  
treatment  
Caution:  
Before plugging the unit in, m ake sure that the voltage rating  
on the back of the AC Adaptor and the supply voltage value  
m atch.  
Press the  
-button for interm ittent operation. The  
nebuliser will not operate if it is not pressed.  
Press the OFF/ON (  
)-button for continuous operation. The  
nebuliser will operate continuously until it is pressed again,  
which stops the unit.  
Rem ove the plug from the m ains socket.  
(Do not pull the power cord. Be sure to pull it out by  
holding the power plug).  
Cleaning  
Notes:  
, Lift the casing up, while pressing the front and rear  
unlock buttons of the AC adaptor. Rem ove the m ain  
unit.  
, Rem ove the cover and the m edication cup holder from  
the m ain unit.  
1. During operation a green power light will illum inate. This  
light m ust be lit for the unit to operate.  
2. The unit is equipped with a heat sensor. If, during operation,  
excessive heat occurs, the unit will autom atically shutoff.  
The unit will operate properly when norm al operating tem -  
perature is reached. Store the unit in a well ventilated loca-  
tion (near an open window for exam ple). Please wait at least  
30 m inutes before restarting inhalation.  
10  
11  
 
English  
Rem ove the nebuliser guide, m outhpiece and aerosol volum e  
adjust knob from the cover.  
Wash the air filter in lukewarm water (40° C or less) or with  
diluted neutral cleanser to reuse. After washing, m ake sure the  
filter is com pletely dry before attaching it to the unit.  
40° C or less  
How to rem ove the delivery rate volum e adjust knob:  
Rotate the knob counterclockwise, adjust the " " m ark  
with the m atching m ark " " of the cover and pull it  
upwards.  
If soiling can no longer be washed out, replace the filter with  
a new air filter (U07-4).  
Other parts  
Pour the m edication out of the cup, clean and sterilise the  
cover, nebuliser guide, m edication cup holder, m outhpiece,  
aerosol volum e adjust knob, rem ovable parts.  
Wash the m edication cup, m edication cup holder, cover, nebuli-  
ser guide, m outhpiece, aerosol delivery rate adjust knob and fan  
in com m ercial cleaning solution (eg. Milton). Do not use soap.  
All parts can be placed in boiling water for 5 to 10 m inutes.  
Do not allow the water to boil dry.  
Unit m aintenance  
Drain water from the water tank. Wipe out rem aining water  
with soft, dry cloth.  
Clean the unit using a cloth m oistened with water and a few  
drops of neutral cleanser, and then wipe dry with a cloth.  
N ever use abrasive m aterial on transducer.  
Be sure that the unit is disconnected from the power source  
before cleaning.  
N ever use benzene, thinner, a volatile chem ical.  
Storage  
Replacing and cleaning of the air filter  
Do not store the unit in direct sunlight or in areas of high  
hum idity. The unit should be kept in a safe place.  
Rem ove the aerosol volum e adjust knob from the cover , ➀  
and rem ove the filter .  
12  
13  
 
English  
Possible problems and how to solve them  
Technical data  
This device fulfills the provisions of the EC directive 93/42/EEC (Medical Device Directive).  
Sym ptom  
Check  
Rem edy  
Operation button Check that the plug of the AC  
Correctly insert it.  
Model:  
OMRON N E U07  
AC 220V – 230V~, 50/60 Hz  
Approx. 20 Watts  
lam p (AC) does  
not light up.  
adaptor is inserted correctly.  
Check that the m ain unit is  
m ounted correctly onto the AC  
adaptor.  
Electrical rating:  
Power consum ption:  
Particle size:  
Correctly m ount it.  
With AC  
adaptor in use.  
3.5 µm (MMD *)  
Check that the level of water in Fill water up to proper level.  
Ultrasonic frequency:  
N ebulisation rate:  
Water tank:  
2.4 MHz  
the water tank is norm al.  
Approx. 1.0 m l/m in.  
Approx. 21 m l  
N o nebulisation  
Check that standard volum e of  
Fill the m edication cup and switch  
on the power by pressing OFF/ON  
the m edication is in the cup.  
Fuse:  
= T 0.1 A  
(
).  
Medication cup:  
Dim ensions:  
m axim um 10 m l  
Low nebulisation Be sure the m edication cup is  
Replace the m edication cup.  
Rem ove the exess m edication.  
Increase delivery rate.  
Approx. 62 m m (W) x 110 m m (D) x 254 m m (H)  
approx. 940 g  
rate  
not broken.  
Check the m edication level in  
cup.  
Check aerosol volum e adjust  
knob.  
Check battery light.  
Check preparation of  
m edication.  
Weight:  
Operating conditions:  
Storage conditions:  
Electrical safety:  
0 °C – 40 °C; 30% RH – 85% RH  
0 °C – 40 °C; 30% RH – 85% RH  
- Class II,  
Go to AC operation.  
Eventually read inform ation sup-  
plied with m edication or consult  
your doctor.  
- Type B  
Sym bol explanation:  
= Manual operation  
AC = Alternating current  
= OFF/ON  
If you find unsolvable problem s please contact your authorised OMRON Custom er Service.  
For address see packaging and/or attached literature.  
Technical data subject to change  
Manufacturer:  
OMRON HEALTHCARE Europe  
* Mass Median Diam eter  
Data independently obtained by:  
Malvern Instrum ents Gm bH, Germ any  
on Malvern Master Sizer System  
14  
15  
 

Nikon Camcorder SB0C0211 6MB07511 02 User Manual
Nikon Projector VT490 User Manual
Nokia TV Receiver 9802 S User Manual
Nortel Networks IP Phone IP Phone 8540 User Manual
Numark Industries Musical Instrument DM 950 User Manual
Nvidia Computer Monitor GD245HQ User Manual
Olympus Printer 330N User Manual
Palm Cell Phone P100EWW User Manual
Panasonic Network Card 2SA1791G User Manual
Panasonic Stereo Equalizer SH MZ1200 User Manual