INTRODUCTION
NOTES ON SAFETY
• The warning signs and the sample icons shown here are listed for you to use the product safely and
correctly as well as to prevent the risk and the damage to you and others from happening.
• The icons and meanings are as follow.
Thank you for purchasing the Omron Automatic Blood Pressure
Monitor with IntelliSense™, HEM-711. Although your blood pressure
fluctuates throughout the day, no self-adjustment is necessary as
the monitor intelligently adjusts to your changing conditions. The
IntelliSense™ monitor uses fuzzy logic intelligence to sense both
your systolic and diastolic blood pressure values, and the result is
personalized inflation with every reading. Arm discomfort and
misreading due to improper cuff inflation are virtually eliminated.
Warning sign
Contents
Indicates matters in which bodily harm or material damage* may arise as a result
of incorrect handling.
Caution
* Material damage refers to a wide range of damage involving your house, household
goods, domestic animals, and pets.
Examples of signs
The
icon indicates prohibitions (what you cannot do).
Matters involving actual prohibitions are indicated by text or pictures in or near
left icon refers to "prohibition to disassemble".
. The
The
icon indicates something compulsory (always follow). Matters involving actual
Your new OMRON monitor uses the oscillometric method of blood
pressure measurement. This means the monitor detects your
blood's movement through your brachial artery and converts the
movements into a digital reading. At the push of a button, the monitor
automatically fills the cuff with air, starts deflation, then displays
your systolic, diastolic, and pulse readings on an extra large display
panel.
compulsory actions are indicated by text or pictures in or near
"general compulsion".
. The left icon refers to
Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
Do not use the unit on infants or persons who cannot express one's
intention.
Please read this instruction manual thoroughly before using
this unit. For specific information on your own blood pressure,
CONTACT YOUR PHYSICIAN.
Do not use the unit for any purpose other than measuring blood
pressure.
Do not use a cellular phone near the unit.
Improper operation may result.
Do not disassemble, repair, or remodel the main unit or the arm cuff of
the blood pressure monitor.
The battery liquid may leak and damage the main unit. Please observe the following points.
• When you are not going to use the unit for a long period of time (approximately three
months or more), take out the batteries.
• Replace the worn batteries with new ones immediately.
• Do not use worn and new batteries together.
• Do not insert the batteries with their polarities in wrong direction.
Do not force to bend the arm cuff or the air tube excessively.
When removing the air tube, pull at the connector.
Do not apply strong shock or drop the main unit.
Do not inflate the arm cuff without it being wrapped on the arm.
5
4
QUICK REFERENCE GUIDE
KNOWYOUR UNIT
1. Avoid eating, smoking, and exercising for at least 30 minutes
before taking a measurement.
Heart Symbol
Systolic Blood
Pressure Display
2. Remove tight-fitting clothing from your upper arm.
R
O
ON/OFF Button
START Button
IT
N
O
M
E
R
U
S
S
E
R
ON/OFF
3. Sit in a chair with your feet flat on the floor and place your arm
on a table so that the cuff is at the same level as your heart.
P
D
O
O
T
U
S
A
LO
Y
E
R
B
M
D
A
E
IC
E
T
E
R
M
A
-711
M
D
TO
U
L H
A
E
O
M
REPLACE
BATTERIES
GRESS
PRO
SYSTOLIC
START
4. Put your arm through the cuff loop making sure that the bottom
edge of the cuff is approximately one-half inch above the elbow
and that the Green Marker on the cuff is above brachial artery.
IC
L
O
T
S
IA
D
PULSE/min
1/2 inch
Memory Call Button
Air Jack
Diastolic Blood Pressure Display /
Pulse Display
Green Marker
Air Plug
(Connects to the air jack)
5. Pull the end of the cuff so that the entire cuff is evenly tightened
around your arm and press the hook material firmly against the
pile side of the cuff.
6. Press the ON/OFF button.
Arm Cuff
7. After the Heart Symbol ( ) appears on the digital panel, press
the Start button, and remain still until the measurement is complete.
Air Tube
Green Marker
8. When measurement is complete, the monitor displays your blood
pressure and pulse rate, and automatically deflates the cuff.
9. Wait 2-3 minutes before taking another blood pressure
measurement. You may require more rest time between readings
depending on your individual physiological characteristics.
Carrying Case
7
6
HOW TO APPLYTHE ARM CUFF
BATTERY INSTALLATION/REPLACEMENT
H
U
L
A
E
D
O
M
1. If the cuff is assembled correctly,
the hook material will be on the
outside of the cuff loop and the
metal D-ring will not touch your
skin.
IC
L
O
T
1. Slide the battery cover off in
the direction of the arrow.
S
Y
S
LIC
TO
IAS
LS
U
P
D
2. Install or replace 4 "AA" size
batteries so the + (positive)
and – (negative) polarities
match the polarities of the
battery compartment as
indicated.
2. If the cuff is not assembled, pass
the end of the cuff furthest from
the tubing through the metal D-
ring to form a loop. The smooth
cloth should be on the inside of
the cuff loop.
Pile Side Material
Sewn Hook Material
3. Replace the battery cover.
3. Sit in a chair with your feet flat
on the floor and place your arm
on a table so that the cuff is at
the same level as your heart.
D-Ring
Notes:
4. Put your arm through the
cuff loop. The bottom of
the cuff should be
If the Low Battery Indicator
(
replace all four batteries.
Long-life alkaline batteries
are recommended.
) appears on the display,
Bottom Edge 1/2 Inch
approximately 1/2"
above the elbow. The
Green Marker on the cuff
should lie over the
brachial artery on the
inside of the arm. Tube
should run down center of arm
approximately even with middle
finger.
Cuff Tube
Remove the batteries if the
monitor will not be used for
an extended period of time.
Green Marker
9
8
HOW TO TAKEA READING
HOW TO APPLYTHE ARM CUFF
5. Pull the cuff so that the top
and bottom edges are
tightened evenly around
your arm.
1. Press the ON/OFF button.
a) All display symbols appear for approximately
one second.
b) When the monitor becomes ready to measure,
the Heart Symbol appears on the display.
2. Press the START button and
remain still.
6. When the cuff is positioned
correctly, press the sewn
hook material FIRMLY
against the pile side of the
cuff.
Initial LCD Test
As the cuff begins to inflate, the monitor
automatically determines your ideal inflation
level. Because this monitor detects the pulse
even during inflation, do not move your arm
and remain still until the entire measurement
completes.
7. Make certain the cuff fits
snugly around your arm.
The cuff should make good
contact with your skin.
3. Inflation stops automatically and
measurement is started.
Ready to Measure
As the cuff slowly deflates, decreasing
numbers appear on the display and the Heart
Symbol ( ) flashes at every heartbeat. In rare
circumstances, a higher inflation may be
necessary. In those cases, the monitor
reinflates the cuff up to 30 mmHg higher than
initial inflation and restarts the measurement.
You should be able to fit
one finger between the
cuff and your arm.
Inflation started.
4. When the measurement is complete,
the arm cuff completely deflates and
your blood pressure and pulse rate
are displayed.
8. Relax your arm and turn
your palm upward.
9. Be sure there are no kinks
in the air tubing.
NOTE: The monitor automatically stores blood
pressure and pulse rate into the memory.
5. Press the ON/OFF button to turn
the monitor off.
NOTE: If you forget to turn the monitor off, it
will automatically shut itself off after five
minutes.
11
10
HOW TO USE THEMEMORY FUNCTION
HOW TO TAKEA READING
This monitor has a memory capable of storing fourteen sets of readings.
Every time you complete the measurement, the monitor automatically
stores blood pressure and pulse rate.
Instructions for special conditions:
• If your systolic pressure is known
to be more than 220 mmHg,
push and hold the START button
until the monitor inflates 30 to 40
mmHg higher than your suspected
systolic pressure.
• Push the
button to recall stored readings while the
Heart Symbol ( ) is displayed.
• When the memory is holding fourteen sets of readings, the
oldest set will be deleted to store a new set.
NOTE: Do not apply more
pressure than necessary.
The monitor will not
inflate above 300 mmHg.
Last reading
• If you want to stop the
measurement, push the ON/OFF
button. The monitor will stop
inflating and start deflating
rapidly, then the monitor will
power off.
START
Value before
the last reading
Value of fourteen
readings before
Continue to press button
• To delete stored readings, press the
button and the
Start button at once, then all the readings will be deleted.
NOTE: You can not delete the stored readings partially.
START
Press these two buttons
at the same time.
Display of memory deletion
13
12
HOW TO USE THE OPTIONAL AC ADAPTER
TROUBLESHOOTING
(ACCESSORY ITEM SOLD SEPARATELY)
To avoid potential damage to the monitor, use only the exclusive
AC adapter (Model HEM-ADPT1) which can be purchased from:
Error Indicator
Cause
Correction
Omron Healthcare, Inc.
300 Lakeview Parkway
Vernon Hills, IL 60061
Toll Free Phone #: 1-800-634-4350
Turn monitor off, wait 2-3 minutes
and take another measurement.
Repeat steps listed under "How to
Take a Reading." If "EE" is displayed
again, take another measurement
by pressing and holding the START
button as instruction page 12 under
"How to Take a Reading", instructions
for special conditions.
Cuff under-inflated.
Movement during
measurement.
To connect AC adapter
Plug
1. Insert the AC adapter plug
into the jack on the side of the
monitor.
Turn the monitor off, carefully read
and repeat steps listed under "How
to Take a Reading."
Cuff over-inflated (more than
300 mmHg).
2. Insert the AC adapter plug
into a 120V AC outlet (50 - 60
cycles).
Jack
Option AC Adapter
To remove the AC adapter
Battery voltage is excessively Replace all four "AA" batteries with
low. new batteries.
To remove the AC adapter, disconnect the adapter plug from the
AC outlet first and then disconnect the cord from the monitor’s
jack.
NOTE: The monitor is designed not to draw power from the batteries
Problem caused by abnormal Contact OMRON Customer Service
memory function. Toll Free 1-800-634-4350
when the AC adapter is in use.
If you continue to have difficulty obtaining a reading, call
OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350.
Please tell the representative you have an OMRON Model HEM-711.
15
14
CAUTION
CARE ANDMAINTENANCE
CAUTION:
To protect your monitor from damage, please AVOID the following:
Changes or modifications not expressly approved by Omron Healthcare, Inc.
could void the user’s warranty.
• Subjecting your monitor and cuff to extreme temperatures,
humidity, moisture, and direct sunlight.
NOTE:
• Folding the cuff and tubing tightly.
• Inflating the monitor over 280 mmHg.
• Disassembling the monitor.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful
interference in a residential installation. The product generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If the product does cause harmful interference to radio or television
reception, which can be determined by turning the product on and off, the user
is encouraged to try to correct the interference by one or more of the following
measures:
• Subjecting the monitor to strong shocks (for example,
dropping the monitor on the floor.)
• Do not clean the monitor with volatile liquids.
THE MONITOR SHOULD BE CLEANED WITH A SOFT,
DRY CLOTH.
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and receiver.
• Connect the product into an outlet on a circuit different from that to
which the receiver is connected.
• Do not clean cuff.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise
emissions from digital apparatus as set out in the interference-causing
equipment standard entitled "Digital Apparatus", ICES-003 of the Canadian
Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques
applicables aux appareils numériques de Classe B prescrites dans la norme
sur le matériel brouilleur: "Appareils Numériques", ICES-003 èdictée par le
ministre des communications.
17
16
FIVE YEAR LIMITED WARRANTY
SPECIFICATIONS
Your HEM-711 IntelliSense™ Automatic Blood Pressure Monitor is
warranted to be free from manufacturing defects for a period of five
years under normal use. The five year warranty excludes the
monitor cuff. The cuff is warranted for a one year period. This
warranty extends only to the original retail purchaser.
Model:
HEM-711
Display:
LCD Digital Display
Measurement Range:
Pressure: 0 to 280 mmHg
Pulse: 40 to 180/minute
Pressure: ±3 mmHg or 2% of reading
Pulse: ±5% of reading
Accuracy/Calibration:
Should repair be needed within the warranty period, ship the unit
prepaid to Omron Healthcare, Inc., 300 Lakeview Parkway,
Vernon Hills, IL 60061, Attn: Repair Dept., together with $5.00 for
return shipping and insurance. Be sure to include the model
number of your unit and your phone number on any
correspondence.
Inflation:
Automatic by electric pump
Active electronic control valve
Deflation:
Rapid Pressure Release:
Pressure Detection:
Measurement Method:
Pulse Wave Detection:
Power Source:
Active electronic control valve
Capacitive pressure sensor
Oscillometric method
We will either repair or replace (at our option) free of charge any
parts necessary to correct defects in the materials or workmanship.
Capacitive pressure sensor
4 "AA" batteries or AC adapter (sold separately)
Approximately 300 uses
Battery Life:
The above warranty is complete and exclusive. The warrantor
expressly disclaims liability for incidental, special, or consequential
damages of any nature. (Some states do not allow the exclusion or
limitation of incidental or consequential damages, so the above
warranty may not apply to you.)
Operating Temperature/Humidity: 50°F to 104°F (10°C to 40°C)
30 to 85% RH maximum
Storage Temperature/Humidity:
–4°F to 140°F (–20°C to 60°C)
10 to 95% RH maximum
Main Unit Weight:
Approximately 12.3 oz (350g) not including batteries
Approximately 4 1/4" (l) x 6 2/3" (w) x 2 3/5" (h)
(109 mm x 170 mm x 66 mm)
Any implied warranties arising by the operation of law shall be
limited in duration to the term of this warranty. (Some states do not
allow limitations on how long an implied warranty lasts, so the
above limitation may not apply to you.)
Main Unit Dimensions:
Cuff Dimensions:
Cuff Circumference:
Accessories:
Approximately 5 1/2" x 19" (140 mm x 480 mm)
Fits arm circumference 9" to 13"
This warranty gives you specific legal rights, and you may have
other rights which vary from state to state. As a condition to
operation of your warranty, the enclosed registration card must be
completed and sent to us within 10 days from the date of purchase.
Arm cuff, carrying case, instruction manual
NOTE: These specifications are subject to change without notice.
*Optional accessories: large cuff (fits arms 13" to 17" in circumference) and small cuff
(fits arms 7” to 9” in circumference) are sold separately.