Omega Engineering Thermometer HH806AW User Manual

comply with emission limits.  
INTRODUCTION  
The instrument is a portable digital thermometer that  
measures external thermocouples of type K, J, R, S, T, E,  
N. The thermocouples types comply with the N.I.S.T. –  
ITS 90 standard reference temperature/voltage tables.  
The thermometer features a dual thermocouple input, an  
adjustable T/C offset and an USB interface for uploading  
data to a PC using optional software and cable.  
Changes or modifications not expressly approved by  
the party responsible for compliance could void the  
user‘s authority to operate the equipment.  
OMEGAnet On-Line Service  
omega.com  
Internet e-mail  
HH806AW  
MULTILOGGER THERMOMETER  
Servicing North America:  
WIRELESS NOTE  
Wireless receiver must keep a distance of at least 30cm  
from the meter and meter to meter distance must be at  
least 20cm.  
M4553/0907  
U.S.A:  
One Omega Drive, Box 4047  
Stamford, CT 06907-0047  
INSTRUCTION  
SHEET  
Shop online at: omega.com e-mail: [email protected]  
For latest product manuals: omegamanual.info  
ISO 9001 Tel: (203) 359-1660  
FAX: (203) 359-7700  
Certified e-mail: [email protected]  
SAFETY INFORMATION  
Canada:  
976 Bergar  
Laval (Quebec) H7L 5A1, Canada  
WARNING  
SPECIFICATIONS  
Tel: (514) 856-6928  
FAX: (514) 856-6886  
To avoid electrical shock, do not use this instrument  
when working voltages at the measurement surface  
over 24V AC or DC.  
ELECTRICAL  
Celsius or Fahrenheit user-selectable  
Measurement Range:  
K-TYPE (0.1°) -200°C to 1372°C (-328°F to 2501°F)  
J-TYPE (0.1°) -210°C to 1200°C (-346°F to 2192°F)  
T-TYPE (0.1°) -200°C to 400°C (-328°F to 752°F)  
E-TYPE (0.1°) -210°C to 1000°C (-346°F to 1832°F)  
For immediate technical or application assistance:  
U.S.A  
and  
Canada:  
Sales Service: 1-800-826-6342/1-800-TC-OMEGA®  
Customer Service: 1-800-622-2378/1-800-622-BEST®  
WARNING  
To avoid damage or burns, do not make temperature  
measurement in microwave ovens.  
Mexico:  
En Espan ol: (001) 203-359-7803  
FAX: (001) 203-359-7807 [email protected]  
R-TYPE (1°)  
S-TYPE (1°)  
0°C to 1767°C (32°F to 3212°F)  
0°C to 1767°C ( 32°F to 3212°F)  
CAUTION  
N-TYPE (0.1°) -50°C to 1300°C ( -58°F to 2372°F)  
*Based on the ITS-90 temperature standard.  
According to temperature standard ITS-90.  
Accuracy:  
Repeated sharp flexing can break the thermocouple  
leads. To prolong lead life, avoid sharp bends in the  
leads, especially near the connector.  
Servicing Europe:  
Benelux: Postbus 8034, 1180 LA Amstelveen, The Netherlands  
Tel: +31 (0)20 3472121 FAX: +31 (0)20 6434643  
Toll Free in Benelux: 0800 0993344  
K/J/T/E-TYPE  
FEDERAL COMMUNICATIONS  
COMMISSION  
±(0.05% rdg + 0.3°C) on -50°C to 1372°C  
±(0.05% rdg + 0.7°C) on -50°C to -210°C  
±(0.05% rdg + 0.6°F) on -58°F to 2501°F  
±(0.05% rdg + 1.4°F) on -58°F to -346°F  
N-TYPE  
±(0.05% rdg + 0.8°C) on -50°C to 0°C  
±(0.05% rdg + 0.4°C) on 0°C to 1300°C  
±(0.05% rdg + 1.6°F) on -58°F to 32°F  
±(0.05% rdg + 0.8°F) on 32°F to 2372°F  
R/S-TYPE  
±(0.05% rdg + 2°C) on 0°C to 1767°C  
±(0.05% rdg + 4°F) on 32°F to 3212°F  
Temperature Coefficient:  
0.1 times the applicable accuracy specification per °C  
from 0°C to 18°C and 28°C to 50°C (32°F to 64°F and  
82°F to122°F).  
Input Protection:  
24V dc or 24V ac rms maximum input voltage on any  
combination of input pins.  
Reading Rate: 2.5 time per second.  
Wireless Features:  
Frequency range: 910~920MHz  
Low current consumption less than 1mA  
The transmitting distance can reach 25M without  
netic interference.  
Czech  
Frystatska 184, 733 01 Karviná, Czech Republic  
Republic: Tel: +420 (0)59 6311899 FAX: +420 (0)59 6311114  
Toll Free: 0800-1-66342 e-mail: [email protected]  
This device complies with Part 15 of the FCC Rules.  
Operation is subject to the following two condi-  
tions:(1) this device may not cause harmful interfer-  
ence, and (2) this device must accept any interference  
received, including interference that may cause unde-  
sired operation.  
France:  
11, rue Jacques Cartier, 78280 Guyancourt, France  
Tel: +33 (0)1 61 37 2900 FAX: +33 (0)1 30 57 5427  
Toll Free in France: 0800 466 342  
Germany/ Daimlerstrasse 26, D-75392 Deckenpfronn, Germany  
Austria:  
Tel: + 49 (0)7056 9398-0 FAX: +49 (0)7056 9398-29  
Toll Free in Germany: 0800 639 7678  
NOTE  
This equipment has been tested and found to comply  
with the limits for a Class B digital device, pursuant to  
Part 15 of the FCC Rules. These limits are designed to  
provide reasonable protection. This equipment gener-  
ates, uses and can radiate radio frequency energy and,  
if not installed and used in accordance with the in-  
structions, may cause harmful interference to radio  
communications. However, there is no guarantee that  
interference will not occur in a particular installation If  
this equipment does cause harmful interference to ra-  
dio or television reception, which can be determined  
by turning the equipment off and on, the user is en-  
couraged to try to correct the interference by one or  
more of the following measures:  
United  
One Omega Drive, River Bend Technology Centre  
Kingdom: Northbank, Irlam, Manchester  
ISO 9002 M44 5BD United Kingdom  
Certified Tel: +44 (0)161 777 6611 FAX: +44 (0)161 777 6622  
Toll Free in United Kingdom: 0800-488-488  
It is the policy of OMEGA Engineering, Inc. to comply with all  
worldwide safety and EMC/EMI regulations that apply. OMEGA  
is constantly pursuing certification of its products to the  
European New Approach Directives. OMEGA will add the CE  
mark to every appropriate device upon certification.  
The information contained in this document is believed to be  
correct, but OMEGA accepts no liability for any errors it  
contains, and reserves the right to alter specifications without  
notice.  
- Reorient or relocate the receiving antenna.  
- Increase the separation between the equipment and  
receiver.  
1
2
These products are not designed for use in, and  
should not be used for, human applications.  
External Connections:  
1. USB Port  
- Connect the equipment into an outlet on a circuit dif-  
ferent from that to which the receiver is connected.  
- Consult the dealer or an experienced radio/TV tech-  
nician for help.  
2. DC power JACK(12V)  
ENVIRONMENTAL  
Ambient Operating Ranges:  
Shielded interface cables must be used in order to  
 
the second display (K/J/T/E/R/S/N). If the inputs of  
the main and second display are the same, then press-  
ing this button will change the thermocouple type for  
both displays.  
Where Do I Find Everything I Need for  
Process Measurement  
(12) “[TIME]” Button (System time setting)  
To set the system time, press the [TIME] button in the  
setup mode. The third display should show the current  
date and time with the year flashing. Enter the new  
value from left to right in the following format  
YY:MM:DD and HH:MM:SS. Press the number but-  
tons printed in white on the overlay to set the system  
date and time. Press the “ENTER” button to confirm.  
Exit this function by pressing ESC button.  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in  
materials and workmanship for period of from date of pur-  
chase. OMEGA’s WARRANTY adds an additional one (1) month grace  
period to the normal to cover handling and  
shipping time. This ensures that OMEGA’s customers receive maximum  
coverage on each product.  
a
and Control?  
OMEGA…Of Course!  
Shop online at omega.com  
4. SETUP MODE  
The following functions can only be use in the setup  
mode.  
If the unit malfunctions, it must be returned to the factory for evaluation.  
OMEGA’s Customer Service Department will issue an Authorized Return  
(AR) number immediately upon phone or written request. Upon examination  
by OMEGA, if the unit is found to be defective, it will be repaired or replaced  
at no charge. OMEGA’s WARRANTY does not apply to defects resulting  
from any action of the purchaser, including but not limited to mishandling,  
improper interfacing, operation outside of design limits, improper repair, or  
unauthorized modification. This WARRANTY is VOID if the unit shows  
evidence of having been tampered with or shows evidence of having been  
damaged as a result of excessive corrosion; or current, heat, moisture or  
vibration; improper specification; misapplication; misuse or other operating  
conditions outside of OMEGA’s control. Components which wear is not  
warranted, include but are not limited to contact points, fuses, and triacs.  
TEMPERATURE  
Thermocouple, RTD & Thermistor Probes,  
Connectors, Panels & Assemblies  
Wire: Thermocouple, RTD & Thermistor  
Calibrators & Ice Point References  
Recorders, Controllers & Process Monitors  
Infrared Pyrometers  
(2)  
Press the LIMITS button to enter the Hi/Lo limit setup  
function. The words “LIMIT” and “Hi” will be flash  
on the LCD along with the previous value for the Hi  
limit. Enter the new Hi limit value using the number  
keys printed in white on the overlay. The resolution of  
Hi/Lo limit setup is 1 count . The “-“ button (same  
button as the ESC) can be used to enter negative val-  
ues. Press the “ENTER” button to confirm the new  
limit. You will then be prompted to enter a new Lo  
limit value. Enter the new value and then press enter  
to finish setup of limits.  
WARNING  
To avoid possible electrical shock, disconnect the  
thermocouple connectors from the thermometer before  
removing the cover.  
PRESSURE, STRAIN AND FORCE  
Transducers & Strain Gages  
Load Cells & Pressure Gages  
Displacement Transducers  
Battery Replacement  
1. Power is supplied by 4pcs 1.5V (AAA SIZE) UM-4  
R03.  
2. The “ ” appears on the LCD display when replace-  
ment is needed. To replace battery remove screw from  
back of meter and lift off the battery cover.  
3. Remove the battery from battery contacts and replace.  
4. When not in use for long periods remove battery.  
5. Do not store in locations with high temperatures, or  
high humidity.  
Instrumentation & Accessories  
FLOW/LEVEL  
Rotameters, Gas Mass Flowmeters & Flow  
Computers  
Air Velocity Indicators  
Turbine/Paddlewheel Systems  
Totalizers & Batch Controllers  
(3) “[INVT]”  
(Interval time setting)  
To setup the time interval for the log function, press  
the [INVT] button. The indicator “INV” will blink on  
the top-right of the LCD and the previous interval is  
displayed. Press the number buttons printed in white  
on the overlay to change the time interval. Setting is  
from left to right of the following format  
(HH:MM:SS). Press the overlay “ENTER” button to  
confirm. To exit this function, press the ESC button.  
HH: interval Hour (0~23)  
pH/CONDUCTIVITY  
pH Electrodes, Testers & Accessories  
Benchtop/Laboratory Meters  
Controllers, Calibrators, Simulators & Pumps  
Industrial pH & Conductivity Equipment  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor  
shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used  
in or with any nuclear installation or activity; or (2) in medical applications or  
used on humans. Should any Product(s) be used in or with any nuclear  
installation or activity, medical application, used on humans, or misused in  
any way, OMEGA assumes no responsibility as set forth in our basic  
Cleaning  
Periodically wipe the case with a damp cloth and deter-  
gent, do not use abrasives or solvents.  
MM: interval Minute (0~59)  
DATA ACQUISITION  
WARRANTY  
/
DISCLAIMER language, and, additionally, purchaser will  
SS: interval Second (0~59)  
MAX: 23:59:59  
MIN: 00:00:01  
Data Acquisition & Engineering Software  
Communications-Based Acquisition Systems  
Plug-in Cards for Apple, IBM & Compatibles  
Datalogging Systems  
indemnify OMEGA and hold OMEGA harmless from any liability or damage  
whatsoever arising out of the use of the Product(s) in such a manner.  
*Software Operation manual is in the Software disk.  
(8) “APO” Button (Auto power off time setting,  
Recorders, Printers & Plotters  
min. 1 minute)  
Direct all warranty and repair requests / inquiries to the OMEGA Customer  
Service Department. BEFORE RETURNING ANY PRODUCT(S) TO  
OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)  
NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN  
ORDER TO AVOID PROCESSING DELAYS). The assigned AR number  
should then be marked on the outside of the return package and on any  
correspondence.  
Use this function to change the time for the auto power  
off (APO) function. Press the APO button in setup  
mode, and the indicator “APO” and the current time  
will flash on the LCD. The default time for APO is 5  
minutes. Press the number button printed in white on  
the overlay to set the APO time. Press the overlay  
“ENTER” button to confirm. To exit this function  
without changing the setting, press the ESC button.  
MAX: 19999 minutes  
HEATERS  
Heating Cable  
Cartridge & Strip Heaters  
Immersion & Band Heaters  
Flexible Heaters  
The purchaser is responsible for shipping charges, freight, insurance and  
proper packaging to prevent breakage in transit.  
Laboratory Heaters  
ENVIRONMENTAL  
FOR  
RETURNS,  
FOR  
please have the following  
information available BEFORE  
contacting OMEGA:  
1. Purchase Order number under  
which the product was  
REPAIRS, consult OMEGA for  
current repair charges. Have the  
following information available  
BEFORE contacting OMEGA:  
1. Purchase Order number to cover  
the COST of the repair,  
2. Model and serial number of the  
product, and  
3. Repair instructions and/or specific  
problems relative to the product.  
MONITORING AND CONTROL  
Metering & Control Instrumentation  
Refractometers  
MIN: 0001 minutes  
(9)[OFS]” Button (Thermocouple offset ad-  
just)  
Pumps & Tubing  
PURCHASED,  
Air, Soil & Water Monitors  
Industrial Water & Wastewater Treatment  
pH, Conductivity & Dissolved Oxygen  
Instruments  
2. Model and serial number of the  
product under warranty, and  
3. Repair instructions and/or specific  
problems relative to the product.  
When the main display input is T1 and a thermocouple  
is connected, the instrument can adjust the offset of the  
thermocouple. The same can be done for T2 when on  
the main display. In the setup mode, press this button  
to enter the thermocouple Offset Setup Function (OFS)  
and the indicator CAL should be displayed on the top-  
right of the LCD and the current setting is also shown.  
Press the number buttons printed in white on the  
overlay to change the offset of the thermocouple. The  
resolution of the setup is 0.1°. Press the “-” button to  
set a negative value. Press the overlay “ENTER” but-  
ton to confirm. MAX: ±1999.9 °C/°F.  
OMEGA’s policy is to make running changes, not model changes, whenever  
an improvement is possible. This affords our customers the latest in tech-  
nology and engineering.  
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.  
© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This  
document may not be copied, photocopied, reproduced, translated, or  
reduced to any electronic medium or machine-readable form, in whole or in  
part, without the prior written consent of OMEGA ENGINEERING, INC.  
Made in Taiwan  
 

MTX Audio Switch ABUS HUB4X8 User Manual
Murphy Network Card 75160 Series User Manual
National Instruments Computer Accessories 321518A 01 User Manual
National Instruments Fan 370753C 01 User Manual
Navman GPS Receiver LA000508 User Manual
NETGEAR Switch GSM7248R User Manual
Onkyo Car Satellite TV System CR 305X User Manual
Panasonic Camcorder AG DVC 15P User Manual
Panasonic Home Security System AW HE120 User Manual
Panasonic Washer CF Y2 User Manual