Omega Engineering Blood Pressure Monitor DPG1000B User Manual

Digital Pressure Gauges  
DPG1000B, DPG1001B, DPG1000BBL,  
DPG1100B, DPG1100BBL  
DPG1000B,  
DPG1000BBL  
Ranges 2000  
and up  
DPG1000B,  
DPG1001B,  
DPG1000BBL  
DPG1100B,  
DPG1100BBL  
DPG1100B,  
DPG1100BBL  
Ranges 2000  
and up  
INSTRUCTION  
SHEET  
M3359/0606  
Shop online at: omega.com e-mail: [email protected]  
For latest product manuals: omegamanual.info  
DPG1000B  
DPG1001B  
DPG1000BBL  
DPG1100B  
DPG1100BBL  
Battery Powered  
Batteries  
Battery Powered, High Accuracy  
Battery Powered w. Backlighting  
Battery Powered, NEMA 4X  
2 AA alkaline  
Battery Life  
1000B, 1001B, 1100B ranges to 1999  
Approx. 2500 hours  
Battery Powered w. Backlighting, NEMA 4X  
Ranges and Resolution  
Resolution is fixed as indicated in table  
Ranges in Bold are standard. Contact Omega Engineering to order other ranges listed below.  
abs: Absolute reference (normally reads atmospheric pressure, reads zero at full vacuum)  
vac: Vacuum gauge, minus sign not used unless specified  
1000B, 1001B, 1100B ranges 2000 and up  
Approx. 2000 hours  
1000BBL, 1100BBL ranges to 1999  
Approx. 180 hrs (backlight on when gauge is on)  
–30.0 inHg/15.0 psig  
–30.0 inHg/100.0 psig  
–30.0 inHg/199.9 psig  
3.00 psig  
120.0 inHg  
199.9 inHg abs  
199.9 inHg  
50.0 oz/in2  
80.0 oz/in2  
240 oz/in2 abs  
240 oz/in2 vac  
240 oz/in2  
1600 mmHg  
760 torr abs  
1600 torr abs  
2100 mmH2O  
3500 mmH2O  
199.9 cmH2O  
350 cmH2O  
1000 cmH2O  
2100 cmH2O  
199.9 mbar  
350 mbar  
1000 mbar abs  
1000 mbar vac  
1000 mbar  
1000 mbar  
2000 mbar abs  
2000 mbar  
4000 mbar  
1.000 bar abs  
1.000 bar vac  
1.000 bar  
35.0 bar  
70.0 bar  
140.0 bar  
199.9 bar  
1.000 kg/cm2 abs  
1.000 kg/cm2 vac  
1.000 kg/cm2  
1.000 kg/cm2  
1.999 kg/cm2 abs  
1.999 kg/cm2  
4.00 kg/cm2  
7.00 kg/cm2 abs  
7.00 kg/cm2  
14.00 kg/cm2  
19.99 kg/cm2  
35.0 kg/cm2  
70.0 kg/cm2  
140.0 kg/cm2  
199.9 kg/cm2  
350 kg/cm2  
1.000 atm abs  
1.000 atm  
1.000 atm  
4.00 atm  
7.00 atm  
14.00 atm  
19.99 atm  
35.0 atm  
70.0 atm  
135.0 atm  
1000BBL, 1100BBL 2000 and up  
Approx. 150 to 1500 hrs depending on backlight  
usage  
5.00 psig  
350 bar  
19.99 kPa  
35.0 kPa  
100.0 kPa abs  
100.0 kPa vac  
100.0 kPa  
100.0 kPa  
199.9 kPa abs  
199.9 kPa  
400 kPa  
700 kPa abs  
700 kPa  
1500 kPa  
1999 kPa  
3500 kPa  
5000 kPa  
3.50 MPa  
7.00 MPa  
14.00 MPa  
19.99 MPa  
35.0 MPa  
1000 g/cm2 abs  
1000 g/cm2  
2100 g/cm2 abs  
2100 g/cm2  
Low Battery Indication  
Low battery symbol on display when batteries  
must be replaced  
15.00 psi abs  
15.00 psig vac  
15.00 psig  
15.00 psig  
30.0 psi abs  
30.0 psig  
60.0 psig  
100.0 psi abs  
100.0 psig  
199.9 psig  
300 psig  
500 psig  
1000 psig  
1999 psig  
3000 psig  
5000 psig  
6.00 inHg  
10.00 inHg  
Dimensions (see drawing on next page)  
DPG1000  
3.5" W x 3.0" H x 2.0" D NEMA 4X housing  
Add approx. 0.75" to height for pressure fitting  
240 oz/in2  
85.0 inH2  
O
140.0 inH2  
O
400 inH2O abs  
400 inH2O vac  
DPG1100  
400 inH2  
400 inH2  
850 inH2  
O
O
O
3.38" W x 2.88" H x 1.65" D housing  
Add approx. 0.75" to height for pressure fitting  
Weight (approximate)  
7.00 ftH2  
12.00 ftH2  
35.0 ftH2  
70.0 ftH2  
140.0 ftH2  
230 ftH2  
O
O
Gauge:  
9 ounces  
O
O
O
O
O
Shipping weight: 1 pound  
Housing Material & Color  
DPG1000  
Extruded aluminum case, light gray epoxy  
powder coated, black ABS/polycarbonate bezel,  
front and rear gaskets, polycarbonate label  
1.000 bar  
1.999 bar abs  
1.999 bar  
4.00 bar  
7.00 bar abs  
7.00 bar  
480 ftH2  
30.0 inHg abs  
30.0 inHg vac  
30.0 inHg  
30.0 inHg  
60.0 inHg abs  
60.0 inHg  
150.0 mmHg  
260 mmHg  
760 mmHg abs  
760 mmHg vac  
760 mmHg  
DPG1100  
199.9 atm  
340 atm  
Light gray ABS/polycarbonate NEMA 4X case,  
rear gasket, polycarbonate label  
14.00 bar  
19.99 bar  
1600 mmHg abs  
Overpressure  
Accuracy (linearity, hysteresis, repeatability)  
0.25% of full scale 1 least significant digit  
DPG1001B: 0.10% of full scale 1 least  
significant digit  
Calibration  
Ranges up to 1999: Front potentiometers, non-  
interactive zero & span, 10% range  
2 times sensor range  
3000 psig range & metric equivalents: 5000 psig  
5000 psig range & metric equivalents: 7500 psig  
Ranges 2000 and up: 112.5% out-of-range  
display: i – – – or i –.–.–.– depending on model  
Ranges 2000 and up: Internal pushbuttons, non-  
interactive zero, midpoint, & span, 10% range.  
Calibration information stored in non-volatile  
memory.  
Display (update rate, type, size)  
3 readings per second nominal display update  
Burst Pressure  
4 times sensor pressure rating, or 10,000 psi,  
whichever is less  
Ranges up to 1999: 31/ digit LCD, 1/2" digit height  
2
Shutoff  
Ranges 2000 and up: 4 digit LCD, 0.4" H digits,  
alphanumeric display of units  
Last character of model number on rear label  
indicates shutoff time. Contact Omega for other  
optional shutoff times.  
Storage Temperature  
–40 to 203°F (–40 to 95°C)  
BBL: Red LED display backlighting  
-5:  
5 minute auto shutoff (standard)  
Controls & Functions  
Operating Temperature  
-NS: No auto shutoff, on/off via front pushbutton  
Ranges up to 1999: Front pushbutton turns  
gauge on/off and starts shutoff timer. Turns on  
backlighting for 1 minute on BBL models.  
–4 to 185°F (–20 to 85°C)  
Connection Size  
Compensated Temperature  
32 to 158°F (0 to 70°C)  
1
/
4" NPT male  
Ranges 2000 and up: Front pushbutton turns  
gauge on/off, starts shutoff timer. Pressing and  
holding front button zeros gauge. The zero  
feature is not used on absolute reference gauges.  
Sensor Material  
All wetted parts are 316 stainless steel  
Temperature Stability  
0.5% span typical, 1% span max. 0 to 82°C  
 
DPG1000B, DPG1001B, DPG1000BBL, DPG1100B, DPG1100BBL  
3
CAUTION:  
Calibration Precautions  
Install or remove gauge using wrench on hex fitting only. Do not attempt to  
tighten by turning housing or any other part of the gauge.  
See label on back of gauge for range!  
Use fittings appropriate for the pressure range of the gauge.  
Do not apply vacuum to gauges not specified for vacuum operation.  
Permanent damage not covered by warranty will result to the sensor.  
DPG1000B, DPG1001B, DPG1000BBL  
Ranges 2000 and up  
DPG1100B, DPG1100BBL  
Ranges 2000 and up  
Due to the hardness of 316 stainless steel, it is recommended that a thread  
sealant be used to ensure leak-free operation.  
NEVER insert objects into the gauge port or blow out with compressed air.  
Permanent damage not covered by warranty will result to the sensor.  
NOTE:  
Calibration Information  
All gauges are factory calibrated on NIST traceable calibration equipment.  
No calibration is required before placing the gauge into service.  
Gauges can be returned to Omega Engineering for factory certified  
recalibration. NIST traceability is available. Gauges can also be recalibrated  
by any metrology lab with pressure calibration equipment at least four times  
more accurate than the gauge.  
DOWN  
UP  
Absolute Reference – These models display atmospheric pressure if the  
gauge port is open to the ambient. It is normal for the reading to constantly  
change in response to atmospheric pressure changes. Vacuum generation  
and atmospheric pressure measurement equipment for accurate calibration  
and thus these are more difficult to calibrate in the field.  
Calibration of Ranges 2000 and up (with 4-digit 2-line display)  
1. Entering the Calibration Mode  
a. Remove the back cover and note the locations of the two internal  
calibration pushbuttons marked UP and DOWN. These buttons are  
disabled unless the gauge is in calibration mode.  
Calibration Preparation – All Models  
a. Calibration should only be performed by qualified individuals using  
appropriate calibration standards and procedures.  
b. With the gauge off, press and hold the DOWN calibration button, and  
also press the front button to power up the gauge in calibration mode.  
2. Calibration Mode Functions  
b. The calibration equipment should be at least four times more accurate  
a. The display first indicates the gauge’s full-scale pressure range, tests all  
display segments, and then indicates CAL to indicate that the gauge is  
in the calibration mode. Release all pushbuttons.  
than the gauge being calibrated.  
c. Remove the screws on the back of the unit and remove cover.  
d. It is good practice to install fresh batteries before calibrating battery-  
b. The display will then indicate the current pressure reading, updating  
approximately 3 times per second. The gauge will remain in the  
calibration mode until powered down or reset manually. While in the  
calibration mode, the shutoff timer, One Touch Zero (gauge reference  
models only), Min/Max (for applicable models) are all disabled, and the  
calibration pushbuttons are active.  
powered gauges.  
e. Allow the gauge to equalize to normal room temperature before  
calibration.  
c. Each press of the UP or DOWN button makes a small correction, which  
may not always be indicated on the digital display. Press and hold the  
pushbutton for one second or longer to make larger continuous  
corrections. The display of the gauge being calibrated is adjusted to  
match the calibrator’s setting or readout.  
d. If the battery pack is unplugged or the power removed during  
calibration, calibration settings will not be saved.  
3. Gauge Reference Gauges (3 Points)  
a. With the gauge port open to atmosphere, the character display will  
alternate between ZERO and CAL. Press the UP and DOWN buttons  
to obtain a display indication of zero.  
DPG1000, DPG1001 - Pull out plastic  
plugs to access zero and span  
potentiometers  
DPG1100 - Unscrew nylon screws to  
access zero and span potentiometers  
b. Apply full-scale pressure. The character display will alternate between  
+SPAN and CAL. Press the UP and DOWN buttons to obtain a  
display indication equal to full-scale pressure.  
Calibration of Ranges up to 1999  
Remove the calibration potentiometer covers on the front of the unit to  
access the zero and span controls.  
c. Apply 50% of full-scale pressure. The character display will alternate  
between +MID and CAL. Press the UP and DOWN buttons to obtain a  
display indication equal to 50% of full-scale pressure.  
Gauge reference units may be re-zeroed without affecting the span  
calibration. The gauge port must be open to the ambient with no pressure or  
vacuum applied. Adjust the Zero control until the gauge reads zero with the  
minus (–) sign occasionally flashing.  
4. Exit Calibration Mode and Verify Calibration  
a. Exit the calibration mode and save the calibration data by pressing and  
holding the front button until the display indicates OFF.  
Span calibration should only be attempted if the user has access to a  
pressure reference of known accuracy. The quality of the calibration is only  
as good as the accuracy of the calibration equipment and ideally should be  
at least four times the gauge accuracy. Zero calibration must be done before  
span calibration. Record readings at three to five points over the range of  
gauge and adjust span control to minimize error and meet specifications.  
b. Verify pressure indications at 0%, 25%, 50%, 75%, and 100% of full  
scale.  
c. Replace the rear cover and screws, taking care not to pinch the power  
leads between the case and the rear cover.  
 
Servicing Europe:  
Benelux:  
Postbus 8034, 1180 LA Amstelveen, The Netherlands  
OMEGAnet® Online Service  
www.omega.com  
Internet e-mail  
Tel: +31 (0)20 3472121  
FAX: +31 (0)20 6434643  
Toll Free in Benelux: 0800 0993344  
Servicing North America:  
One Omega Drive, Box 4047  
Stamford CT 06907-0047  
Tel: (203) 359-1660  
Czech Republic: Frystatska 184, 733 01 Karviná, Czech Republic  
USA:  
ISO 9001 Certified  
Tel: +420 (0)59 6311899  
Toll Free: 0800-1-66342  
FAX: +420 (0)59 6311114  
FAX: (203) 359-7700  
France:  
11, rue Jacques Cartier, 78280 Guyancourt, France  
Tel: +33 (0)1 61 37 2900  
Toll Free in France: 0800 466 342  
FAX: +33 (0)1 30 57 5427  
Canada:  
976 Bergar  
Laval (Quebec) H7L 5A1, Canada  
Tel: (514) 856-6928  
FAX: (514) 856-6886  
Germany/Austria: Daimlerstrasse 26, D-75392 Deckenpfronn, Germany  
Tel: +49 (0)7056 9398-0  
FAX: +49 (0)7056 9398-29  
Toll Free in Germany: 0800 639 7678  
For immediate technical or application assistance:  
®
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA  
®
United Kingdom: One Omega Drive, River Bend Technology Centre  
Customer Service: 1-800-622-2378 / 1-800-622-BEST  
®
ISO 9002 Certified  
Northbank, Irlam, Manchester  
M44 5BD United Kingdom  
Tel: +44 (0)161 777 6611  
Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN  
TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA  
FAX: +44 (0)161 777 6622  
Mexico:  
En Espan˜ol: (001) 203-359-7803  
FAX: (001) 203-359-7807  
Toll Free in United Kingdom: 0800-488-488  
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply. OMEGA is constantly pursuing certification of its products to the  
European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.  
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc. accepts no liability for any errors it contains, and reserves the right to alter specifications without  
notice. WARNING: These products are not designed for use in, and should not be used for, human applications.  
WARRANTY/DISCLAIMER  
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase.  
OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and  
shipping time. This ensures that OMEGA’s customers receive maximum coverage on each product.  
If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will issue an Authorized Return (AR)  
number immediately upon phone or written request. Upon examination by OMEGA, if the unit is found to be defective, it will be repaired or replaced  
at no charge. OMEGA’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling,  
improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows  
evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or  
vibration; improper specification; misapplication; misuse or other operating conditions outside of OMEGA’s control. Components which wear are not  
warranted, including but not limited to contact points, fuses, and triacs.  
OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither assumes responsibility for any  
omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided  
by OMEGA, either verbal or written. OMEGA warrants only that the parts manufactured by it will be as specified and free of defects.  
OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED, EXCEPT THAT OF  
TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE  
ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of  
OMEGA with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not  
exceed the purchase price of the component upon which liability is based. In no event shall OMEGA be liable for consequential, incidental or  
special damages.  
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in  
or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear  
installation or activity, medical application, used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic  
WARRANTY/ DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless from any liability or damage  
whatsoever arising out of the use of the Product(s) in such a manner.  
RETURN REQUESTS / INQUIRIES  
Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO  
OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER  
TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any  
correspondence.  
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit.  
PATENT NOTICE: U. S. Pat. No. 6,074,089; 5,465,838 / Canada 2,228,333; 2,116,055 / UK GB 2,321,712 / Holland 1008153 / Israel 123052 / France  
2 762 908 / EPO 0614194. Other patents pending.  
FOR WARRANTY RETURNS, please have the following information  
available BEFORE contacting OMEGA:  
FOR NON-WARRANTY REPAIRS, consult OMEGA for current repair charges.  
Have the following information available BEFORE contacting OMEGA:  
1. Purchase Order number under which the product was PURCHASED,  
2. Model and serial number of the product under warranty, and  
3. Repair instructions and/or specific problems relative to the product.  
1. Purchase Order number to cover the COST of the repair,  
2. Model and serial number of the product, and  
3. Repair instructions and/or specific problems relative to the product.  
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and  
engineering.  
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.  
© Copyright 2004 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic  
medium or machine-readable form, in whole or in part, without the prior written consent of OMEGA ENGINEERING, INC.  
M3359/0606  
 

Multiquip Portable Generator GAC 6HZR User Manual
Multiquip Welding System GLW 180H User Manual
Multi Tech Systems Network Card CC1600 Series User Manual
Murphy Switch LLS User Manual
NetComm Network Card NCT1000 User Manual
Nuance comm Computer Accessories 2889A G00 180 User Manual
Olympus DVR DS 10 User Manual
Onkyo Radio T 411RDS T 409 User Manual
Optiquest Computer Monitor VS12105 User Manual
Parasound Stereo Amplifier HCA 1201A User Manual