Milwaukee Doll MR100ATC User Manual

MR100ATC - Salinity Refractometer  
Specifications  
Scale Range Resolution  
Parts Per Thousand 0 to 100 0/00 1 0/00  
Specific Gravity 1.000 to 1.070  
Dimensions 7.6 x 1.5 x 1.5" (194 x 38 x 38mm)  
Weight Approx. 8.0 oz. (227g)  
Shadow-line Æ  
Operation  
The instrument measures the refractive index of the sample and displays the result in parts  
per thousand ( 0/00 ) and specific gravity.  
1. Zero Adjustment  
Put distilled water on the prism; enough to cover the entire prism. Close the cover plate and rotate  
the adjusting screw so that the light/dark boundary line (known as the shadow-line) evens up with  
the zero line. After the zero adjustment, clean the prism with soft cloth.  
2. Sample Preparation and Reading  
To take a reading, place a few drops of a sample liquid on the measurement prism. Ensure that  
enough solution is added to the prism in order to cover the entire prism. Close the prism so that  
the liquid spreads across the entire surface of the prism without air bubbles or dry spots (see  
diagrams below). Allow the sample to remain on the prism for approximately 30 seconds. While  
holding the instrument under a light source, look through the eyepiece. The salinity concentration  
is determined by the intersection of the boundary of the light and dark fields (known as the  
shadow-line) on the printed scale. The left side of the scale indicates the specific gravity and the  
right side parts per thousand. If the scale appears out of focus, the eyepiece may be adjusted by  
rotating the knurled portion. The instrument also features an eye guard to prevent stray light from  
entering the eyepiece and causing reflections. It may be necessary to adjust the position of the  
light source to maximize the contrast of the shadow-line. Under normal conditions, optimal  
contrast is obtained by holding the instrument underneath and perpendicular to a light source.  
(florescent light is not recommended) - Once a reading has been taken, wipe dry with a clean  
cloth (do not wash or rinse) and place the instrument in the supplied plastic case. Store the  
instrument in a safe, dry environment.  
 

Miele Clothes Dryer T 9446 C User Manual
Miele Cooktop KM5732 User Manual
Miele Oven H4892BP2 User Manual
MTD Lawn Mower 111 200 User Manual
Olympus Battery Charger BU 300 User Manual
Optimus Portable CD Player CD 3327 User Manual
Oreck Vacuum Cleaner XL5 105HH User Manual
Panasonic DVD Player DVD CV35 User Manual
Panasonic MP3 Player CQ C5302W User Manual
Patton electronic Network Router 6081RC User Manual